1. Confirm whether the power cable is properly installed and properly connected according to the specifications.
Two, confirm whether the water storage is enough and whether the lid of the water tank is properly covered and clean.
3. Confirm whether the wet bulb gauze is placed correctly, and place it at the front end of the humidity sensor.
Iv. Whether the wet bulb gauze has been wetted (there is water in the water supply tank, if the water level is normal and the gauze cannot be wetted, please replace the gauze immediately, usually about every three months)
5. Whether the water level of the water supply tank is normal.
6. Check whether the water level of the humidifier is normal. Check the water level 3 to 5 minutes after the humidifier is supplied with water.
Seven. Whether the water is clean during the humidification period, the humidifier should be drained regularly to keep the interior clean.
8. Check that the drainage pipe is connected normally, without blockage of foreign matter, and can not be higher than the bottom of the box.
Nine, check and confirm that the overtemperature protection setting is 5℃-10℃ higher than the current experimental conditions.
(Note: gauze should be removed for temperature test only. Gauze will embrittlement after running at high temperature above 85℃. Gauze should be replaced before the next equipment test operation.)
Through the high and low temperature wet heat test box boot matters to know, I hope to help customers and friends do a good job in equipment maintenance. If the equipment has problems, we should contact hongzhan after-sales service team for inspection and maintenance in order to ensure the normal test of customers, reasonable arrangement of production, prolong the life of equipment, and better carry forward the national sustainable development and green energy saving requirements.